AIRIS
II GANTRY AND PATIENT HANDLING SYSTEM
The gantry
design, incorporating a vertical field magnet, provides
a spacious appearance and a comfortable environment for
the patient during scanning.
·
Magnet opening: 43cm
vertical height
·
Off-center post
placement for unrestricted access from all four sides
·
Extra-wide (80cm)
table with 500 lb. weight capacity
·
Power driven table
movement in all three directions; vertical, horizontal
and lateral
·
Manual table-top
emergency release
·
Laser light
localization in three planes for accurate patient
positioning
·
Start scan function
available at front, sides and rear of magnet gantry
·
Patient intercom
·
Padded table-top
AIRIS
II GRADIENT SYSTEM
The
HITACHI-designed and built gradient system provides
sub-millimeter in-plane spatial resolution and
thin-slice imaging. The pulse-gradient driver’s
state-of-the-art digital electronic technology provides
high reliability and efficient operation,
·
Gradient strength:
l5 mT/m, maximum
·
Slew rate: 20 T/m/s,
maximum
·
Completely air
cooled gradients
·
Shielding: Passive,
to neutralize the effects of eddy currents
·
Eddy current
compensation: 1/10th factor, maximum
·
Soft Sound gradients
offer extremely low acoustic noise for added patient
comfort
AIRIS
II RF SYSTEM
The RF
system provides for highly reliable transmission of RF
and reception of MR signals. An ultra low-noise
pre-amplifier and pre-matched RF coil electronics
provide for maximum performance.
·
4 channel
QUAD transmitter for uniform excitation
·
5.0 kw transrnitter
power amplifier to accommodate even the largest patients
·
0.3 dB ultra-low
noise pre-amplifier for high SNR
·
Actively decoupled
RF receiver Coils
·
Linear and
quadrature detection
·
Auto tuning
AIRIS
II RF RECEIVER COIL SET
Open QUAD
Head Coil
This spacious quadrature coil design provides high
signal-to-noise, which is especially useful for
efficient routine brain imaging, as well as
high-resolution, thin-slice imaging.
QUAD
Extremity Coil
This effective quadrature coil uses a high performance
quadrature/solenoid design to provide high
signal-to-noise, which is especially useful for high
resolution imaging of detailed knee anatomy.
Large
Extremity Coil
A high sensitivity design to
optimize imaging of shoulders, large knees, thighs and
pediatric heads.
Volumetric
Neck and Extremity Coil
A high sensitivity solenoid
coil for neck and extremity imaging applications. This
volumetric coil provides high signal-to-noise and a
uniform field of view within the coil’s 20 cm diameter.
Medium
QUAD Flexible Spine and Body Notch Coil
This medium spine and body coil (120cm in circumference)
combines quadrature RF technology with solenoid coil
design for high quality spine and body imaging. The wrap
around features offers uniform signal intensity within
the selected field-of-view, along with an extended
longitudinal coverage for comprehensive lumbar or
thoracic spine imaging. Patient comfort and fit is
enhanced with the notched, flexible design.
Large QUAD Flexible Spine and Body Notch Coil
This large spine and body coil (150cm in circumference)
combines quadrature RP technology with solenoid coil
design for high quality spine and body imaging. The wrap
around features offers uniform signal intensity within
the selected field-of-view, along with an extended
longitudinal coverage for comprehensive lumbar or
thoracic spine imaging. Patient comfort and fit is
enhanced with the notched, flexible design.
AIRIS
II COMPUTER SYSTEM
The
Hitachi designed computer system was developed for use
by a single operator. The state-of-the-art performance
and graphic user interface (GUI) allows for rapid system
operation.
·
Host CPU is a 64
bit, RISC-based UNIX processor
·
512 Mbytes memory
·
Additional
processors for pulse sequence control, data acquisition,
image reconstruction and image processing
·
<0.ls reconstruction
time (256 x 256)
·
Easy-to-use
Openwindows-based GUI
·
Multi-tasking
operating environment
AIRIS
II OPERATOR CONSOLE
The
control console provides for straightforward operation
through the use of optimized ergonomic component design.
·
21” high resolution
color monitor designed and manufactured by HITACHI
·
Mouse input for
rapid interaction
·
Keyboard input for
direct alpha-numeric response
·
Control panel for
patient intercom and emergency stop
AIRIS
II OPERATING SOFTWARE
The
operating software configuration allows for easy access
to all operator functions including scanning,
reconstruction. display/analysis,
and archiving.
·
Scan Task Card
o
Scan protocol
selection
o
Slice and
pre-saturation positioning
o
Display of
reconstructed images
·
Viewport
Tool
Maneuvering of images such as
window/level, cine, ROI and measurement
· 2D
Processing Task Card
Performs 2D post-processing functions including
Adaptive Reconstruction and
edge enhancement
· 3D
Processing Task Card
Performs 3D post-processing functions including MPR and
MIP
· Analysis
Task Card
Supports quantitative analysis features
· Maintenance
Task Card
Supports specialized procedures to insure high image
quality
AIRIS
II IMAGING CAPABILITIES
The image
acquisition software provides multiple slice imaging,
with the flexibility of user-selectable imaging
parameters.
· Automatic
Per-Patient Prescan
Set RF power
Set frequency
Set receiver gain
Set receiver tuning
Perform auto-shim
· Image
Plane Selection
Transverse, sagittal and
coronal
Single and double oblique angles
Multiple slice, multiple angle
· Acquisition
Features
2D multislice with contiguous or gapped slices (up to
256 slices)
3D single or multiple slabs (up to 32)
User variable image matrix (64-512 x 64-512)
Maximum of 4 echoes for multiecho sequences
20 slice thickness of 2mm -100mm
3D slice thickness of 0.5mm - 5mm
· Standard
Pulse Sequences
2D Spin Echo (SE)
2D Inversion Recovery (IR)
2D STIR
2D FLAIR
2D Gradient Echo (GE)
2D Dual Slice
3D Spin Echo (SE)
3D Inversion Recovery (IR)
3D Gradient Echo (GE)
· Steady
State Pulse Sequences
2D SARGETM (SG)
2D PP-Spoiled SARGE (RSSG)
2D Time-Reversed SARGE (TRSG)
3D SARGE (SG)
3D RF-Spoiled SARGE (RSSG)
3D Time-Reversed SARGE (TRSG)
· MR
Angiography
2D Time-of-Flight (TOP) 2D TOF with RF Spoiled SARGE (RSSG)
3D Time-of-Flight (TOF)
3D TOF with RF Spoiled SARGE (RSSG)
Sloped Slab Profile (SSP) for reducing
through-plane flow saturation
2D TOF with walking pre-saturation to highlight either
arterial or venous flow
Compatible with:
Gradient moment nulling for flow compensation
Overlapping 3D slabs
Presaturation (up to 6)
Breath-hold
Cardiac gating
Peripheral pulse gating
Swap frequency/phase direction
2D slice thickness of 2mm -1 00mm
3D slice thickness of 0.5mm - 5mm
Variable flip angle
Half echo
Featuring:
High Resolution/High Definition (HR/HO) reconstruction
for increased visualization
of small vascular structures.
Maximum Intensity Projection (MIP) processes 2D and 3D
acquired images
to calculate projection images with batch or user
defined rotations. Includes
selective ROI.
· Fast
Spin Echo
2D Fast Spin Echo (FSE)
2D Fast Inversion Recovery (FIR)
3D Fast Spin Echo (P52)
3D Fast Inversion Recovery (FIR)
FSE STIR
FSE FLAIR
Single contrast (T1, T2 or proton density)
Dual contrast (T2 and proton density)
Compatible with:
Auto bandwidth
Gradient moment nulling for flow compensation
User variable image matrix (128-1024 x 128-512)
2D slice thickness of 2mm -100mm
3D slice thickness of 0.8mm - 5mm
Presaturation
Rectangular FOV
Off-center FOV
Featuring:
User variable echo factor
User variable interecho
spacing
· Protocol
Enhancements (compatible with various pulse sequences)
Variable bandwidth
Auto bandwidth
No frequency/phase wrap
Swap frequency/phase direction
Presaturation (up to 6)
Rectangular FOV
Off-center FOV
Gradient moment nulling for flow compensation
Cardiac gating
Peripheral pulse gating
Variable flip angle
Half scan
Half echo
AIRIS
II IMAGE PROCESSING/DISPLAY/ANALYSIS
Rapid
image handling and precise image manipulation is enabled
with the high performance AIRIS II computer system.
·
Image Display
o
Display matrix: 1280
x 1024
o
Variable window
level and width adjustment
o
User pre-set window
level and width
o
Image acquisition
annotation
o
Multiple image
display
o
User selectable
image layout
o
Reverse Image
·
Image Processing
o
Magnify
image
o
Rotate image
o
Edge enhancement
with other post processing algorithms
o
Background
noise reduction (BNR)
o
Cine and multi-image
cine
·
Image Analysis
o
Distance measure
o
Angle measure
o
Line profile plot
o
Region of interest
(ROI)
o
Histogram
· Adaptive
Reconstruction
For enhanced image signal to noise. Adaptive
Reconstruction can be programmed
as part of image acquisition process and is user
selectable from a predefined set,
or can be executed as a 2D
post-processing task
· Maximum
Intensity Projection (MIP)
Processes 2D and 3D acquired images to calculate
projection images with batch
or user defined rotations. Used, for example, with MRCP
and MRA.
· Expanded
Multi-Tasking
The AIRIS II multi-tasking provides simultaneous
scan and reconstruction
capabilities. This feature increases Throughput during
multiple acquisitions by
eliminating the time interval associated with
reconstruction. Other multi-tasking
features include image review,
prescribe ahead scan set-up, archive, filming MPR,
MP and network transfers (optional).
AIRIS
II ARCHIVE/INTERFACE
Integral
to the AIRIS II operating system is the support of short
and long term archive
and digital interface for filming.
· Image
Archive
Magnetic disk: 4 GB storage capacity for over 9800
images (256x256)
Optical disk: 2.6 GB re-writeable storage capacity for
over 17,000
images (256x256)
· Interface
Digital output and host control of laser imager